ClearanceBrief
Recall Z-1228-2014 ยท DSQ Ventricular (Assist) Bypass

Thoratec Corporation

HeartMate II System Controller Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own. It is sometimes called a a heart pump or VAD. HeartMate II is a small implantable LVAD. HeartMate II attaches to the heart and is designed to assist or take over the pumping function of the patients left ventri

DateMar 21, 2014 (initiated Mar 4, 2014)
StatusTerminated
ReasonDiscovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Root causeDevice Design
ActionThoratec sent an Urgent Medical Device Correction letter dated March 4, 2014 to all affected customers. The letter was sent via e-mail. All distributors outside of the USA and Canada to whom affected product has been distributed have been notified of this field correction notice and have been provided copies of the Customer Letter to forward to their customers. The letter identified the affected product , problem and actions to be taken. Customers were requested to complete and sign the attached Acknowledgement Form and return it to Thoratec via fax (925-847-8571) or e-mail a scanned copy
Quantity4120
DistributionWorldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Le
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.