Recall Z-1640-2014 ยท KOI Stimulator, Nerve, Peripheral, Electric

GE Healthcare, LLC

GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.

DateMay 21, 2014 (initiated Apr 9, 2014)
StatusTerminated
ReasonWhen the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a
Root causeSoftware design
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter (Ref 36099) dated April 9, 2014 to affected customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information (contact Technical Support at 1-800-558-7044 or your local Service Representative). UPDATE: On May 11, 2015 a second GE Healthcare letter FMI 36099-2 was sent to OUS consignees. The updated report was for additional affected modules due to FRU installation.
Quantity1,794 (471 units US, 1,323 units OUS)
DistributionUS Nationwide Distribution in the states of: AR, CA, CO,DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MI, MN, MO, NC, NV, NY, OH, OK, TN, TX, UT, WA, WI, WY. OUS: ARGENTINA. AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EG
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051635DATEX-OHMEDA S/5 NEUROMUSCULAR TRANSMISSION MODULE, E-NMT AND ACCESSORIES
GE Healthcare, Jul 15, 2005
K955026AS/3 NMT MODULE
Datex Medical Instrumentation, Inc., Apr 28, 1997