Recall Z-1553-2014 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Siemens Healthcare Diagnostics

ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP, Catalog # 078-A011R03 - ADVIA Centaur XP refurb.

DateMay 13, 2014 (initiated Feb 18, 2014)
StatusTerminated
ReasonInstrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
Root causeDevice Design
ActionOn February 18, 2014 Siemens Healthcare Diagnostics distributed Urgent Medical Device Correction notices to their customers via courier service (Urgent Medical Device Correction/). Upon receipt of the communication, customers are asked to follow "Actions to be Taken by the Customer". If the cover is unable to stay partially raised, immediately contact your Siemens technical support representative to arrange for a service. Complete and return the Field Correction Effectiveness Check attached to this letter within thirty (30) days by faxing to (302) 631-7597. Customers with questions should
Quantity5,900 units total (1,679 domestically & 4,221 internationally)
DistributionNationwide and internationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K971418CHIRON DIAGNOSTICS AUTOMATED CHEMILUMINESCENCE SYSTEM (ASC) CENTAUR ANALYZER
Chiron Diagnostics Corp., Jul 17, 1997