ClearanceBrief
Recall Z-1218-2014 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Sarns Malleable Dual-stage Venous Return Cannulae, 28/38 Fr with 1/2" connector, 14.5" (37 cm) long The Sarns Malleable Dual-stage Venous Return Catheters and Sarns Dual-stage Venous Return Cannula are indicated for single-tube venous drainage from the right atrium and vena cava during cardiopulmonary bypass surgery. These devices are indicated for up to 6 hours of use.

DateMar 19, 2014 (initiated Mar 3, 2014)
StatusTerminated
ReasonDuring in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.
Root causeDevice Design
ActionTerumo sent an Urgent Medical Device Recall letter dated March 3, 2014, to all affected consignees. The letter included the reason for removal, potential hazards, and customer instructions. Terumo CVS advised users to stop using the affected devices and return any unused product on hand. Questions or concerns were directed to Terumo CVS Customer Service: 1-800-521-2818 M-F, 8am-6pm. For questions regarding this recall call 734-741-6173.
DistributionWorldwide Distribution - USA including CA, CT, FL, GA, IL, IN, LA, MI, MN, MO, NC, NY, PA, TN, TX, and Internationallly to United Arab Emirates (UAE), Taiwan, Thailand, Malaysia, Belgium, Japan, Mexico, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K920902MODIFICATION OF SARNS TWO STAGE VENOUS CANNULA
3M Health Care, Sarns, May 11, 1993