ClearanceBrief
Recall Z-1483-2014 ยท FRN Pump, Infusion

Hospira Inc.

Hospira Acclaim Encore Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.

DateMay 1, 2014 (initiated Oct 31, 2013)
StatusTerminated
ReasonBroken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.
Root causeComponent design/selection
ActionURGENT MEDICAL DEVICE CORRECTION letters, dated October 31, 2013, were sent to all direct accounts. The letters identified the affected product, as well as the reason for recall, risk to health, and instructions for device users. Users are to inspect each Hospira/Abbott Acclaim Encore infusion pump for door handle cracks prior to programming a therapy by following the provided instructions. The letter also included instructions for recipients to ensure that all potential users in the facility are made aware of the safety notification and the recommended actions; and complete the attached re
Quantity13,218 infusion pumps total
DistributionNationwide Distribution -- AK, AL, AR, AZ, CA, CO, CR, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011096ABBOTT ACCLAIM ENCORE, MODEL 12237
Abbott Laboratories, May 1, 2001