Recall Z-1143-2014 ยท CBK Ventilator, Continuous, Facility Use

Phillips Respironics, Inc.

Trilogy Ventilators The Respironics Trilogy 100 system provides continuous or intermittent ventilator support for the care of individuals who require mechanical ventilation. Trilogy 100 is intended for pediatric through adult patients weighing at least 5 kg (11 lbs.). The device is intended to be used in home, institution/hospital, and portable applications such as wheelchairs and gurneys, and may

DateMar 6, 2014 (initiated Feb 11, 2014)
StatusTerminated
ReasonInternal testing revealed a potentially defective component on the power management board that could affect the function of certain Trilogy Ventilators resulting in loss of therapy to patients.
Root causeComponent design/selection
ActionPhilips sent an Urgent - Medical Device Recall letter on February 11, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the information contained in the Notice with all members of their staff who need to be aware of the contents of the communication. For further information or support concerning the issue, customers were instructed to contact Philips Respironics at 877-387-3311. For questions regarding this recall call 724-733-0200.
Quantity572
DistributionWorldwide Distribution - USA (nationwide) and the countries of France, Great Britian, China, India, Italy, Korea, Kuwait, Nehterlands, and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083526TRILOGY 100 VENTILATOR
Respironics Inc., Sleep & Home Respiratory Group, Mar 13, 2009
K093416TRILOGY 200 VENTILATOR
Respironics, Inc., Jan 29, 2010
K093905TRILOGY 202 VENTILATOR
Respironics, Inc., May 12, 2010