Recall Z-1145-2014 ยท LYB

China Daheng Group Inc.

Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical powered. Surgical powered Laser instruments.

DateMar 14, 2014 (initiated Jul 15, 2013)
StatusTerminated
ReasonFailure to provide calibration procedures to purchasers, as required by 21 CFR 1040.11(a)(2). For example, the user manual did not contain calibration procedures for Denlase or Penlase Lasers
Root causeRadiation Control for Health and Safety Act
ActionChina Daheng Group, lnc.(CDH}, ensures tha they will execute seriously their plans for the repair of their electronics products Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System. China Daheng Group, lnc.(CDH}, will send their updated 238 pes of Diode Laser Therapy System operation manual and 200 pes of Dental soft Tissue Laser brochure to their Distributor. They will pay all of their expense including manufacture cost and freight cost. China Daheng Group, lnc.(CDH}, will remedy these defects through changing the manuals and brochures for their distri
Quantity238
DistributionCA, KS, IL, PA, NJ, WI, WY, KY, FL and NY India, Indonesia, Turkey, Denmark, and France,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.