Recall Z-1145-2014 ยท LYB
China Daheng Group Inc.
Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System surgical powered. Surgical powered Laser instruments.
| Date | Mar 14, 2014 (initiated Jul 15, 2013) |
| Status | Terminated |
| Reason | Failure to provide calibration procedures to purchasers, as required by 21 CFR 1040.11(a)(2). For example, the user manual did not contain calibration procedures for Denlase or Penlase Lasers |
| Root cause | Radiation Control for Health and Safety Act |
| Action | China Daheng Group, lnc.(CDH}, ensures tha they will execute seriously their plans for the repair of their electronics products Denlase: Diode Laser Therapy System and Penlase: Dental Laser Therapy System. China Daheng Group, lnc.(CDH}, will send their updated 238 pes of Diode Laser Therapy System operation manual and 200 pes of Dental soft Tissue Laser brochure to their Distributor. They will pay all of their expense including manufacture cost and freight cost. China Daheng Group, lnc.(CDH}, will remedy these defects through changing the manuals and brochures for their distri |
| Quantity | 238 |
| Distribution | CA, KS, IL, PA, NJ, WI, WY, KY, FL and NY India, Indonesia, Turkey, Denmark, and France, |