Recall Z-1308-2014 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RaySearch Ray Station, Radiation Therapy Treatment Planning System, Stand-alone software

DateMar 31, 2014 (initiated Feb 6, 2014)
StatusTerminated
ReasonRaySearch Laboratories has recalled RaySearch Ray Station 4.0 software due to clinical dose calculation errors during radiation therapy.
Root causeSoftware Design Change
ActionOn February 6, 2014, RaySearch Laboratories distributed Urgent Medical Device Correction notices to their customers via email regarding the affected product. To determine if the version you are using is affected, bring up the About box in the RayStation application and check if the number reported is "4.0.2.9". If so, the notice applies to your version. Customers are advised to inform all planning staff and all users about the instructions listed on the notice. Customers are advised to inspect their system and identify all installed units with the referenced software version numbers listed
Quantity19 units (16 units domestically & 3 units internationally)
DistributionOH, MD, NJ, WA, IL, MA, FL, Spain, United Kingdom, Italy,