Recall Z-1307-2014 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation Radiation Therapy Treatment Planning System, models 2.5, 3.0, 3.5 and 4.0; a software system designed for treatment planning and analysis of radiation therapy.

DateMar 28, 2014 (initiated Feb 6, 2014)
StatusTerminated
ReasonPotential for dose errors due to software program errors.
Root causeSoftware Design Change
ActionRaySearch Laboratories sent an Field Safety Notice, Medical Device Correction letter dated February 4, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. ACTIONS TO BE TAKEN BY THE USER A workaround that eliminates the described error is to select the discard rotations option in the Image Registration Module before creating a rigid registration intended to be used for hybrid deformable registration. Another option is to use the ROI based deformable registration when applicable, as this functionality is not affecte
Quantity151 units Total (88 units domestically & 63 units internationally)
DistributionWorldwide Distribution - US including the states of NY, OH, MD, CT, CO, TN, IL, AZ, MS, NJ, PA, DE, FL, CA and WA., and the countries of New Zealand, Australia, Republic of Korea, China, Italy, Germany, Norway, Belgium, Switzerland, United Kingdom and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100552RAYSTATION VERSION 1.0
Raysearch Laboratories AB, Mar 12, 2010
K120387RAYSTATION
Raysearch Laboratories AB, Apr 6, 2012
K130617RAYSTATION
Raysearch Laboratories AB, May 10, 2013