ClearanceBrief
Recall Z-1157-2014 ยท OMP Negative Pressure Wound Therapy Powered Suction Pump

Smith & Nephew Inc.

Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

DateMar 10, 2014 (initiated Jan 9, 2014)
StatusTerminated
ReasonReports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
Root causePackaging change control
ActionA product bulletin titled "Soft Port Enhancements" dated March 14, 2013, was released for external distribution describing the changes to the Soft Port products, e.g. increasing the Soft Port orifices size and adding a second slit to the release paper covering the port adhesive, and stating the purpose of the modifications to the device. These product bulletin was provided to sales representatives for distribution to customers requesting information regarding the design enhancement made to the Soft Port product range. The bulletin was not distributed to all the customers. To ensure that the
Quantity162,629 invididual units
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110647RENASYS FOAM NPWT WOUND DRESSING KITS WITH SUCTION PAD; RENASYS GAUZE NPWT WOUND DRESSING KITS WITH SUCTION PAD; RENASYS
Smith & Nephew, Inc., Jun 22, 2011