Recall Z-1704-2014 ยท FRN Pump, Infusion
Hospira Inc.
Plum XL3 Multi line Infusion System, Micro Macro Plum XL3 Multi line Infusion System, and Micro Macro Plum XL Infusion System. Infusion pumps for intravenous. arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products to patients in hospital and home care environments.
| Date | Jun 3, 2014 (initiated Jan 24, 2014) |
| Status | Terminated |
| Reason | There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion |
| Root cause | Component design/selection |
| Action | Hospira sent an URGENT MEDICAL DEVICE RECALL letter dated January 24, 2014, to all affected customers via traceable UPS. The letters included instructions for recipients to: inspect the door roller assemblies; if a door roller or pin appears loose, broken or missing, remove the device from clinical service until the door assembly is replaced with a corrected door assembly and contact Hospira at 800-441-4100 (Monday - Friday, 8:00 AM - 5:00 PM CST) to report the issue. All devices requiring repair will be serviced and replaced with the existing design door and/or pin. Hospira is in the proces |
| Quantity | List Number 11781: 1,828 pumps; List Number 11845: 13,700 pumps; List Number 11846: 22,671 pumps |
| Distribution | Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K952799 | ABBOTT XL3M INFUSION PUMP Abbott Laboratories, Jan 18, 1996 |