Recall Z-1193-2014 ยท JIT Calibrator, Secondary

Ortho-Clinical Diagnostics

VITROS Immunodiagnostic Products 25-OH Vitamin D Total Calibrators; REF/CATALOG # 684 4055, IVD --- COMMON/USUAL NAME: VITROS Vitamin D Calibrators -- Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK --- For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic Systems, and the VITROS 5600 Integrated System for the quantitative measurement of

DateMar 12, 2014 (initiated Jan 15, 2014)
StatusTerminated
ReasonOrtho Clinical Diagnostics has determined that the VITROS Immunodiagnostic Products 25-OH Vitamin D Total Calibrators, Lots 0091 through 0235, used with VITROS Immunodiagnostic Products 25-OH Vitamin D Total Reagent Packs, may generate negatively biased results. As a result, users must immediately discontinue using all current lots of VITROS 25-OH Vitamin D Total Calibrators and Reagent Packs.
Root causeComponent design/selection
ActionOn 1/22/2014, Customer Letters were sent via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees to inform them of the issue and advise them to discontinue use of the affected lot(s) and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. On 1/22/2014, Distributor Letters were sent via US Postal Service Priority Mail and/or FedEx overnight courier to US distributor consignees and instructed them to notify their consignees of the issue and actions.
Quantity1554 units domestically
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121608VITROS IMMUNODIAGNOSTIC PRODUCTS 25-OH VITAMIN D TOTAL REAGENT PACK, VITAMIN D TOTAL CALIBRATORS
Ortho-Clinical Diagnostics, Inc., Feb 15, 2013