Recall Z-1120-2014 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems, Inc.

HeartStart MRx Defibrillator/Monitor M3538A Lithium Ion Batteries Revision J The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the

DateFeb 26, 2014 (initiated Feb 18, 2014)
StatusTerminated
ReasonPhilips HeartStart MRx Monitor/Defibrillator, when operating on battery power only, may experience an unexpected shutdown if exposed to elevated levels of electromagnetic interference from RF energy source
Root causeDevice Design
ActionThe firm, Philips Healthcare, sent an "URGENT- Medical Device Correction" letter dated February 2014 and issued a Field Safety Notice on February 18/ 2014 to its customers. The letter and field safety notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace the affected batteries free of charge. Philips instructed customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice. Fixed, portabl
Quantity43,792 batteries
DistributionWorldwide Distribution: US (nationwide) and Internationally to: ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGUIM, BOLIVIA, BRAZIL, CANADA, CHINA, COLOMBIA, CONGO, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
Philips Medical Systems, Sep 8, 2005