Recall Z-0996-2014 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Medtronic Inc. Cardiac Rhythm Disease Management
Medtronic Select 3D Arterial Cannulae. Vented, model 78622 (22 Fr.). Non-Vented Cannulae, model 78722 (22 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic. The Cannula consists of a curved, angled or beveled tip with or without flange that is permanently attached to a molded, clear, flexible PVC tapered body or wirebound PVC body. The cannula body features a tip orientation line to in
| Date | Feb 20, 2014 (initiated Jan 30, 2014) |
| Status | Terminated |
| Reason | Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar. |
| Root cause | Under Investigation by firm |
| Action | The firm, Medtronic, sent an "Urgent Medical Device Recall" letter dated January, 2014 via UPS on January 30, 2014 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately identify and quarantine all listed product in your inventory; return all listed product to Medtronic and contact Customer Service at 800 854-3570 and reference RADAR #167826 to initiate a return and credit of product; complete and return the attached Customer Confirmation Certificate via fax to Medtronic at 651-367-0612 to the attention of Customer |
| Quantity | 19483 total (7669 US, 11814 OUS) |
| Distribution | Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paragu |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K013013 | MEDTRONIC DLP ARTERIAL CANNULA WITH 3D TIP - 22 FR. Medtronic Cardiac Surgical Products, Dec 4, 2001 |