Recall Z-1128-2014 ยท EHD Unit, X-Ray, Extraoral With Timer

Carestream Health Inc.

Model Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- Carestream Health, Inc. Made in France -- --- COMMON/USUAL NAME: Dental X-Ray Systems -- CLASSIFICATION NAME: Unit, X-Ray, Extraoral with timer --- There are four major components: x-ray tube, x-ray generator, collimator and image receptor. These systems are permanently installed diagnostic denta

DateMar 4, 2014 (initiated Jan 24, 2014)
StatusTerminated
ReasonCarestream received problem reports relating to apparent early failure of Kodak 2X00 Intraoral X-ray device arms leading to arm breakage.
Root causeComponent design/selection
ActionUrgent Medical Device Recall Letters (dated 1/24/2014) were sent to the consignees on 1/24/2014 via Certified Mail, Return Receipt Requested.
QuantityDomestic: 347 units (total for both)
DistributionWorldwide distribution. USA (nationwide) and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060292KODAK MODELS 2100 AND 2200
Kodak-Trophy, Mar 3, 2006