Recall Z-0845-2014 ยท NCX System, Hypothermia, Intravenous, Cooling

Philips Healthcare

InnerCool RTx Endovascular System, Model Numbers: 861470 (120V), 861472 (240V). Product Usage: The InnerCool RTx device is a thermal regulating system intended to induce, maintain and reverse mild hypothermia, to achieve and/or maintain normothermia, and for use in fever reduction. K080908

DateJan 27, 2014 (initiated Jan 13, 2014)
StatusTerminated
ReasonPhillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.
Root causeNonconforming Material/Component
ActionPhilips Healthcare Recall Notice for this proactive performance correction was provided verbally to hospital biomedical engineers. Philips product support engineering scheduled and conducted the update on customer devices to the most current MCB configuration.
Quantity50
DistributionWorldwide Distribution - US (Nationwide) and Internationally to: Austria, France, Germany The Netherlands, Spain, Sweden, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080908INNERCOOL THERAPIES, INC., RAPIDBLUE SYSTEM
Innercool Therapies, Inc.- A Delaware Corporation, Oct 17, 2008