ClearanceBrief
Recall Z-1098-2014 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

Optiflux F160NRe High Flux Dialyzers; Product Number: 0500316E. Single Use acute or chronic hemanalysis treatment.

DateFeb 26, 2014 (initiated Feb 4, 2014)
StatusTerminated
ReasonDialysate Port Leak During Priming of the Dialyzers
Root causeUnder Investigation by firm
ActionFresenius Medical issued Notice to Clinics and Clinics with Patients Letters Dated - 1st Mailing - February 4, 2014 and Patients Letters Dated - February 5, 2014 - these letters contained the correct information.Corrected letters to Clinics and Clinics with Patients issued 2/14/14 via certified mail with return receipt and fax-back form . Users are advised the leak may be stopped by tightening the dialysate port cap more securely. Questions associated with this product notification to your Customer Service Team at 1-800-323-5188.
Quantity3,460,672 units
DistributionWorldwide distribution: USA (nationwide) and the country of: Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002761FRESNIUS HEMOFLOW F7NR, F50NR, F70NR HEMODIALYZERS
Fresenius Medical Care North America, Dec 4, 2000