Recall Z-0900-2014 ยท ITX Transducer, Ultrasonic, Diagnostic

Boston Scientific Corporation

Atlantis PV, 8.5F, 15 MHz Peripheral Imaging Catheter: Sterile, R; Sterilized using irradiation. Product Usage: This catheter is a 15 MHz ultrasound imaging catheter intended to operate with an IVUS instrument for diagnostic imaging. It is used with an 8.5F-introducer sheath and a 0.035 guidewire.

DateJan 30, 2014 (initiated Jan 6, 2014)
StatusTerminated
ReasonReports of physicians experiencing inability to pass a 0.035" guidewire through the guidewire lumen and exit port at the (proximal) Y-manifold of the imaging catheter.
Root causeEquipment maintenance
ActionBoston Scientific sent an Urgent Medical Device Recall Removal letter dated January 14, 2014 to all affected customers by overnight mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete an Account Reply verification Tracking Form in order to received a Return Goods Authorization number for product return. For questions contact your local Sales Representative.
Quantity1068 single units
DistributionUS Nationwide Distribution and Mexico