ClearanceBrief
Recall Z-0846-2014 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Focus Diagnostics Inc

Simplexa Flu A/B & RSV Direct assay, Model MOL2650. The Focus Diagnostics Simplexa Flu A/B & RSV Direct assay is intended for use on the 3M Integrated Cycler instrument for the in vitro qualitative detection and differentiation of influenza A virus, influenza B virus, and respiratory syncytial virus (RSV) RNA in nasopharyngeal swabs (NPS) from human patients with signs and symptoms of respiratory

DateJan 28, 2014 (initiated Jan 10, 2014)
StatusTerminated
ReasonFocus Diagnostics is initiating an urgent safety notice correction for Simplexa Flu A/B & RSV Direct assay (MOL2650) because Focus Diagnostics has received some customer complaints of Simplexa Flu A/B & RSV lots due to sporadic false positive RSV signals, which may result in a higher RSV false positive rate.
Root causePending
ActionFocus Diagnostics sent an Urgent Safety Notice Correction letter dated January 10, 2014, via email and a hard copy letter by mail. The letter informed customers of the correction, product description with codes, product intended use, issue, risk to health, background, recommendation, actions by customers, and contact information. Customers were instructed to check their inventory for the affected product, if customers detect any positive RSV results, confirm the positive result by an alternate method. Customers were asked to complete the "URGENT: SAFETY NOTICE Acknowledgement and Receipt Form
Quantity1,052 kits
DistributionWorldwide Distribution - US (Nationwide) including states of: NY, MA, NJ, OH, CA, HI, CO, MN, FL, PA, IA, WI, AZ, TX, TN, IL, SD, MO, NM, AR, WV, MI and Internationally to: Sweden, Israel, Canada, and Slovenia. .
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120413SIMPLEXA FLU A/B & RSV DIRECT, SIMPLEXA FLU A/B & RSV POSITIVE CONTROL PACK
Focus Diagnostics, Inc., Jul 13, 2012