Recall Z-0908-2014 ยท LWR Heart-Valve, Non-Allograft Tissue

Medtronic Cardiovascular Surgery-the Heart Valve Division

Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses.

DateFeb 3, 2014 (initiated Jan 9, 2014)
StatusTerminated
ReasonMedtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring post implant.
Root causeDevice Design
ActionMedtronic sent an Customer Notification letter dated January 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Medtronic recommends surgeons continue to remain diligent in selecting the appropriate size of Mosaic Aortic Bioprosthesis. Adoption of the updated Mosaic Aortic Bioprosthesis sizing chart (iEOA chart) and Mosaic Obturators/Sizers will facilitate optimal aortic valve size selection. Your Medtronic Field Representative will be contacting you about receiving a new Mosaic Aortic Obturator/Sizer set and
Quantity8276 units
DistributionWorldwide Distribution - USA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.