Recall Z-0908-2014 ยท LWR Heart-Valve, Non-Allograft Tissue
Medtronic Cardiovascular Surgery-the Heart Valve Division
Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses.
| Date | Feb 3, 2014 (initiated Jan 9, 2014) |
| Status | Terminated |
| Reason | Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring post implant. |
| Root cause | Device Design |
| Action | Medtronic sent an Customer Notification letter dated January 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Medtronic recommends surgeons continue to remain diligent in selecting the appropriate size of Mosaic Aortic Bioprosthesis. Adoption of the updated Mosaic Aortic Bioprosthesis sizing chart (iEOA chart) and Mosaic Obturators/Sizers will facilitate optimal aortic valve size selection. Your Medtronic Field Representative will be contacting you about receiving a new Mosaic Aortic Obturator/Sizer set and |
| Quantity | 8276 units |
| Distribution | Worldwide Distribution - USA (nationwide) |
| FDA record | Open on fda.gov |