Recall Z-1076-2014 ยท JIX Calibrator, Multi-Analyte Mixture

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.

DateFeb 24, 2014 (initiated Jan 6, 2014)
StatusTerminated
ReasonComplaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) and imprecision during PCHE calibration with the Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL).
Root causeNonconforming Material/Component
ActionA recall notification letter, dated January 2014, was sent to Users.
Quantity3496
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061923DIMENSION VISTA ENZYME 1 CALIBRATOR (ENZ 1 CAL - KC310)
Dade Behring, Inc., Aug 25, 2006