ClearanceBrief
Recall Z-0915-2014 ยท BTM Ventilator, Emergency, Manual (Resuscitator)

GE Healthcare

Stand-Alone Resuscitation Unit, Bag and Mask, Upgrade Kits, part number M1139612, labeled in part ***GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com*** Provides the basic equipment required for pulmonary resuscitation of infant.s

DateFeb 3, 2014 (initiated Dec 12, 2013)
StatusTerminated
ReasonMedical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.
Root causeProcess control
ActionGE Healthcare notified end users with a letter flagged as Urgent Medical Device Correction letter on 12/12/2013. The notification advised that GE Healthcare has become aware of a potential safety issue associated with the loosening of the blender knob shaft on all Resuscitation units. The firm advised that accounts ensure that all potential users in their facility are made aware of this safety notification and the recommended actions described in the notification: 1. Perform blender accuracy check as outlined in Section 2.5-2.5.1 of the Service Manual to verify that the O2 concentration set
Quantity48 units
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Uruguay, Venezuela, Albania, Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Canada, Chile, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guatem
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070247GIRAFFE AND PANDA BAG AND MASK RESUSCITATION SYSTEM
Ohmeda Medical, Apr 20, 2007