ClearanceBrief
Recall Z-1517-2014 ยท JHR Radioimmunoassay, Immunoglobulins (D, E)

Siemens Healthcare Diagnostics

IMMULITE /IMMULITE 1000 Systems TIE Total IgE, REF LKIE1, SMN 10380867, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.

DateApr 25, 2014 (initiated Dec 20, 2013)
StatusTerminated
ReasonSiemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.
Root causeUnder Investigation by firm
ActionUrgent Medical Device Correction Letters (dated 12/20/13) and effectiveness check response forms were sent to customers on 12/20/13 via FedEx to inform them of the issue. Customers were requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within thirty (30) days. Field service personnel were sent a support document describing the issue and instructing them how to deal with customer questions.
Quantity5953 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K931703IMMULITE TOTAL IGE
Diagnostic Products Corp., Aug 16, 1993
K960067IMMILITE TOTAL IGE
Diagnostic Products Corp., Feb 14, 1996