Recall Z-0866-2014 ยท HXX Screwdriver

Medtronic Sofamor Danek USA Inc

SET SCREW RETAINING DRIVER, REF 9010000820, QTY: 1 EA, Medtronic Sofamor Danek, USA, Inc. Screwdriver - orthopedic manual surgical instrument.

DateJan 30, 2014 (initiated Jan 3, 2014)
StatusTerminated
ReasonThe retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.
Root causeReprocessing Controls
ActionThe recall notices were mailed directly to Risk Managers and Sales Representatives, with Recall Questionnaires and Confirmation forms included, on 01/03/2014.
Quantity7 units
DistributionUS Distribution in the states of CA, NC, TX, IN, NJ, and MI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.