Recall Z-0844-2014 ยท KPR System, X-Ray, Stationary

Carestream Health, Inc.

KODAK DirectView DR 7500 Dual Detector System, MODEL DR 7500, Manufactured by Carestream Health, Inc. Made in U.S.A.

DateJan 27, 2014 (initiated Dec 26, 2013)
StatusTerminated
ReasonCarestream Health, Inc. has recalled DR 7500 Dual Detector System utilizing Version 5.X Software due to a possible patient image display error.
Root causeSoftware Manufacturing/Software Deployment
ActionOn December 26, 2013 Carestream Health, Inc. distributed Urgent Medical Device Recall notification letters dated December 19, 2013 to their customers via Fed Ex. Carestream Health, Inc. recommends that once the daily detector calibration process is initiated on the DR 7500 System running Version 5.X software it should not be interrupted. The calibration process should be completed only when patients are not being imaged. A Carestream Health Service representative will contact all customers to install a software modiciation to Version 5.X software which will not allow the cancellation of det
Quantity164 units Total (42 domestically & 122 internationally)
Distributionworldwide Distribution - US (Nationwide) including states of: IN, PA, MD, MI, NY, GA, MN, OK, VA, TX, WI, CA, IL, TN, WA, NC, SC, MA, NJ, and KY; and Internationally to: Canada, New Zealand, South Korea, Australia, Hong Kong, Saudia Arabia, Guinea, Kuwait, Israel, Oman, South Africa, China, UK, Swed
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051258KODAK DIRECTVIEW DR 7500 SYSTEM, MODEL 8791345
Eastman Kodak Company, Jun 1, 2005