Recall Z-1085-2014 ยท IYE Accelerator, Linear, Medical
Elekta, Inc.
Agility. Version 3.1 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.
| Date | Mar 6, 2014 (initiated Nov 25, 2013) |
| Status | Terminated |
| Reason | During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value. |
| Root cause | Software design |
| Action | Corrective Action #1 - Release of Important Field Safety Notice 200 01 502 043 - "Recommended quality assurance after a calibration procedure" Released 25 November, 2013. Corrective Action #2 - Permanent Solution - All users of Integrity 3.1 software will be informed or a mandatory upgrade to Integrity 3.2 upon its release which will prevent calibration from being inverted. Target Release for Integrity 3.2 will be available at the end of February 2014. |
| Quantity | 134 |
| Distribution | Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CT, FL, GA, IL, KY, LA, MD, MA, MI, MO, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, and WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K123808 | AGILITY Elekta , Ltd., Apr 4, 2013 |