Recall Z-0779-2014 ยท ERL Drill, Surgical, Ent (Electric Or Pneumatic) Including Handpiece

The Anspach Effort, Inc.

Anspach MIA 16 (16cm Minimally Invasive Attachment), Rx Only MIA16 attachments are intended for use with Motor Systems for cutting and shaping bones, including the spine and cranium.

DateJan 16, 2014 (initiated May 6, 2013)
StatusTerminated
ReasonProduct did not reach expected sterility assurance level
Root causeUnknown/Undetermined by firm
ActionThe firm, Aspach, sent an "URGENT: MEDICAL DEVICE LABELING CORRECTION" letter dated May 13, 2013 to its consignees/customers. The letter described the product, problem and actions to be taken. The consignees/ customers were instructed that Sterrad 100S system should not be used on the MIA 16 attachment (the information only applies to only Sterrad sterilization of this devices and not to any other attachments.), and complete and return the Return Receipt Requested Customer Reply form via fax to: 1-800-327-6661 or email: ProductSupportTeam@synthes.com. If you distribute any of the products t
Quantity183
DistributionNationwide Distribution: FL, IL, CA. IN, MI, CA, NJ, TX, NE, NC, OR, PA, TN, DC, UT, CO, MO, KY, MD, WV, NY, and MN.