Recall Z-1024-2014 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer, Inc.

TRILOGY ACET SHELL 36MM O TRILOGY ACET SHELL 38MM O TRILOGY ACET SHELL 40MM O TRILOGY ACET SHELL 42MM O TRILOGY ACET SHELL 44MM O TRILOGY ACET SHELL 46MM O TRILOGY ACET SHELL 48MM O TRILOGY ACET SHELL 50MM O TRILOGY ACET SHELL 52MM O TRILOGY ACET SHELL 54MM O TRILOGY ACET SHELL 56MM O TRILOGY ACET SHELL 58MM O TRILOGY ACET SHELL 60MM O TRILOGY ACET SHELL 62MM O TRILOGY ACET SHELL 64MM O TRILOGY AC

DateFeb 20, 2014 (initiated Oct 11, 2013)
StatusTerminated
ReasonZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These p
Root causePackaging
ActionDistributors received a letter identifying the issue and their responsibilities dated November 4, 2013. These responsibilities include quarantining the affected fielded inventory in their possession, providing hospital risk managers with the recall notification letter and retrieving affected fielded inventory and direct sales from the hospitals.
Quantity7,044,680 total
DistributionWorldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934765TRILOGY ACETABULAR SYSTEM
Zimmer, Inc., Apr 29, 1994
K953490TRILOGY ACETABULAR SHELL & LINERS 36MM
Zimmer, Inc., Oct 20, 1995
K954698TRILOGY ACETABULAR COMPONENT 7MM OFFSET
Zimmer, Inc., Jan 17, 1996