Recall Z-0989-2014 ยท JXG Shunt, Central Nervous System And Components

Aesculap, Inc.

Aesculap proSA Adjustment Disc Size Large (L) The Miethke proSA Adjustment shunt system is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

DateFeb 19, 2014 (initiated Dec 23, 2013)
StatusTerminated
ReasonThe accuracy is out of specification for the Aesculap proSA Adjustment Disc Size L.
Root causeEmployee error
ActionAesculap sent an Important Correction and Removal Notification dated December 23, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers.. Yom"Aesculap Sales Representative will contact you to schedule a time to remove the FV795T from Set FV792T. Please use FV790T in the absence of the FV795T to ensure the correct adjustment has been made. Please complete the attached Inventory Sheet which is necessary to comply with FDA regulations. When completing the inventory sheet, please fill in the quantity being returned. If
Quantity79
DistributionWorldwide Distribution - US Distribution including the states of AZ, MD, MA, WI, MI, MN, DE, KS, and FL., and the countries of Australia, Germany, Great Britain, Norway and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120559AESCULAP-MEITHKE PROSA ADJUSTABLE SHUNT SYSTEM
Aesculap, Inc., May 24, 2012