Recall Z-0932-2014 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476)

DateFeb 4, 2014 (initiated Dec 4, 2013)
StatusTerminated
ReasonThe PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.
Root causeSoftware design
ActionOn 12/4/2013, the firm sent Field Safety Noticies to their customers.
Quantity386 units
DistributionProduct was shipped to the following states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, NY , OH, OR, PA, TN, TX, VA, VT, WA & WI. Product was also shipped to the following countries: Australia, Belgium, Brazil, Bulgaria, Canada, China, Cr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052640GEMINI RAPTOR
Philips Medical Systems (Cleveland), Inc., Oct 7, 2005
K081135GEMINI CONDOR
Philips Medical Systems (Cleveland), Inc., Apr 29, 2008