Recall Z-0932-2014 ยท KPS System, Tomography, Computed, Emission
Philips Medical Systems (Cleveland) Inc
Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476)
| Date | Feb 4, 2014 (initiated Dec 4, 2013) |
| Status | Terminated |
| Reason | The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study. |
| Root cause | Software design |
| Action | On 12/4/2013, the firm sent Field Safety Noticies to their customers. |
| Quantity | 386 units |
| Distribution | Product was shipped to the following states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, NY , OH, OR, PA, TN, TX, VA, VT, WA & WI. Product was also shipped to the following countries: Australia, Belgium, Brazil, Bulgaria, Canada, China, Cr |