Recall Z-0990-2014 ยท FKB Connector, Blood Tubing, Infusion T

Perrigo Diabetes Care

Orbit soft Infusion sets, 9 mm soft cannula, P10 (42") tubing; Single-Use Set Orbit Infusion Sets are sterile, non-pyrogenic, single-use devices intended to be used for subcutaneous delivery of medication , for example insulin, from an external infusion pump.

DateFeb 19, 2014 (initiated Dec 20, 2013)
StatusTerminated
ReasonPerrigo was notified by our supplier, Ypsomed AG, that they are recalling these batches due to being manufactured with a non-conforming female luer-lock connector. The luer-lock connectors are incompatible with standard male luer-lock connectors for use in connection to compatible infusion pumps. Perrigo became was notified of a quality complaint related to this issue on November 20, 2013 from a
Root causeNonconforming Material/Component
ActionPerrigo Diabetes Care sent a recall notice dated December 16, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate the affected productt and request return of all undistributed/unused sets from their customer warehouses to Omni Manufacturing Services S.A. de C.V., Parque Industrial Queretaro, La Noria Lt. 13. Mz 3 No. 125, Santo Rosa Jauregui, MX-76220 Santiago de Queretaro. Customers were instructed to distribute the attached "Urgent Field Safety Notice" to all users or patients w
Quantity7924 units
DistributionNationwide Distribution including GA & TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.