Recall Z-0716-2014 ยท HRS Plate, Fixation, Bone

Biomet, Inc.

Locking Cortical Screw 3.5mmX38mm REF 8161-35-038

DateJan 10, 2014 (initiated Dec 19, 2013)
StatusTerminated
Reason3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw.
Root causeLabeling mix-ups
ActionURGENT MEDICAL DEVICE RECALL notices were sent to all consignees on December 19, 2013. Distributors who have affected hospitals in their regions were notified and provided with a copy of the notice that went to each hospital. Consignees are being contacted on an individual basis in an attempt to reconcile the product 100%. Fax back forms were provided to all consignees in order for Biomet to track returns. Non-responders will be contacted on an individual basis.
Quantity52
DistributionUS Distribution including states of: AZ, CA, GA, IL, KY, MI, NC, NH, NY, OH, PA, SD, TX, UT, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072083SMALL FRAGMENT LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Oct 25, 2007
K072832ANTEROLATERAL AND MEDIAL LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Dec 14, 2007
K082300ANATOMIC LOCKING PLATING SYSTEM
DePuy Orthopaedics, Inc., Feb 18, 2009