ClearanceBrief
Recall Z-0930-2014 ยท DIP Enzyme Immunoassay, Diphenylhydantoin

Siemens Healthcare Diagnostics, Inc

ADVIA Centaur Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems

DateFeb 4, 2014 (initiated Dec 18, 2013)
StatusTerminated
ReasonADVIA Centaur Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.
Root causeUnder Investigation by firm
ActionSiemens Healthcare Diagnostics Customer Notification notice was sent to affected Siemens Healthcare Diagnostic customers in the United States on December 18, 2013.. The notifications inform customers that Siemens Healthcare Diagnostics received customer feedback regarding the performance of the ADVIA Centara Systems Phenytoin assay, indicating that the assay was recovering higher relative to the performance of other assays in proficiency studies. Siemens internal investigation confirmed the positive bias for Calibrator N lot CN 57 in the ADVIA Centaur Systems Phenytoin assay. This lot has expi
Quantity1294 (144 units US)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011393RANDOX PHENYTOIN
Randox Laboratories, Ltd., Dec 4, 2001