ClearanceBrief
Recall Z-0647-2014 ยท CAW Generator, Oxygen, Portable

On Site Gas Systems Inc.

On Site Gas Systems POGS 33C Portable Oxygen Generation System

DateJan 8, 2014 (initiated Dec 19, 2013)
StatusTerminated
ReasonHeat exchanger hose becomes weak and may fail.
Root causeDevice Design
ActionOn Site Gas Systems issued letter on 12/19/13 and stated replacement parts to retrofit the blue Federal Heater hose with stainless steel will be provided and the letter included the Technical Bulletin TB 12-19-13-1 .
Quantity198 units
DistributionUS Distribution: PA only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.