Recall Z-0785-2014 ยท IZJ Camera, X-Ray, Fluorographic, Cine Or Spot

Philips Medical Systems (Cleveland) Inc

ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems. Distributed by Philips Medical Systems, Cleveland, OH

DateJan 17, 2014 (initiated Dec 4, 2013)
StatusTerminated
ReasonThe recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.
Root causeNonconforming Material/Component
ActionOn October 30, 2013 & December 4, 2013, the firm sent Urgent Medical Device Recall letters to their customers.
Quantity255 units
DistributionWorldwide Distribution: US (nationwide) including states of: AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K913471CIRRUS GAMMA CAMERA
Adac Laboratories, Oct 23, 1991