Recall Z-0804-2014 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare, LLC

Video Monitor Suspensions that may be on the following Fluoroscopic and Radiographic X-Ray systems: Precision 500D, Legacy, Prestilix, Monitrol, RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII Product Usage: The systems are intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations on both adult and pediatric subjects. The systmes ar

DateJan 21, 2014 (initiated Nov 8, 2013)
StatusTerminated
ReasonMissing screws in the overhead Video Monitor Suspension. GE discovered that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were missing since installation of the system. GE is not aware of any fall of the overhead monitor suspension.
Root causeEmployee error
ActionGE Healthcare Systems sent an " Urgent Medial Device Correction" letter dated November 7, 2013. The letter was addressed to Hospital Administrators / Risk Mangers, Managers of Radiology /Cardiology and Radiologists / Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information (USA 800-437-1171, Japan 0120-055-919, For other countries, please contact your local GE Healthcare Service Representative).
Quantity5651
DistributionWorldwide Distribution - US Nationwide including Washington D.C., Guam, and Puerto Rico. Outside the US to include: Afghanistan, Algeria, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina faso, Cameroon, Canada, Chile, China, Colo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001
K081091PRECISION 500D R&F X-RAY SYSTEM
GE Healthcare, Jun 11, 2008
K801320SFX-90 TABLE/SPOTFILMER W/FLUORICON L200
General Electric Co., Jul 21, 1980
K862120SCX SYSTEM
General Electric Co., Jun 16, 1986
K936180TC+ SYSTEM (TILTING C-ARM SYSTEM)
GE Medical Systems, Mar 16, 1994
K943805PRESTIGE REMOTE R & F SYSTEM MODELS H, VH & VHD
General Electric Co., Sep 23, 1994