Recall Z-0747-2014 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare, LLC

GE Healthcare Carescape Patient Data Module The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes.

DateJan 14, 2014 (initiated Nov 27, 2013)
StatusTerminated
ReasonGE Healthcare has recently become aware of a potential safety issue due to intermittent failure of the spring loaded mounting latch of PDMs produced from April 2007 through December 2012.
Root causeDevice Design
ActionThe firm, GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 26, 2013 to its consignees/customers. The letter was addressed to Healthcare Administrator / Risk Manger, Chief of Nursing and Director of Biomedical Engineering. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The consignees/customers were instructed to please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. GE Healthcare will provide a new latch at no
Quantity43,033 (25,681 USA; 17,352 OUS)
DistributionWorldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, ID, IL, IN, IA, KS,KY, LA, ME, MD, MA, MI, MN, MS, NO, NE, NV, NJ, NM, NY, NC, OH, OK,OR, PA, RI, SC, TN, TX, UT,VT, VA, WA, WI, and WY. Internationally to: Argentina, Australia, Austria, Bangladesh, Bel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071073SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE
Ge Medical Systems Information Technologies, May 11, 2007