Recall Z-0607-2014 ยท NEP System/Device, Pharmacy Compounding

Baxter Healthcare Corp.

IntelliFill i.v. Pharmacy Compounding Device (PCD). Intended to fill syringes from 0.5mL up to 11.5mL using mixing instructions from its formulary to fill the orders.

DateDec 29, 2013 (initiated Nov 8, 2013)
StatusTerminated
ReasonFollowing a retrospective review for Baxter Daytona Complaint CR-954, it was identified that the IntelliFill i.v. Pharmacy Compounding Device (PCD) (product code: IFIV) has the potential to continue to run and fill syringes with the HEPA filtration disabled.
Root causeNonconforming Material/Component
ActionA Service Bulletin was sent to all IntelliFill i. v. customers February 22, 20 I 0 by USPS mail. The Service Bulletin informs customers that IntelliFill i.v. operators are to be trained to listen for the running HEPA filtration fan motor and observe/feel for airflow coming from the automation deck. The field service engineer dispatched to the site replaced the damaged sensor on 11-Feb-2010 and a Service Bulletin was sent to all customers regarding the function and use of the HEPA filter on 22-Feb-2010. A new HEPA sensor was designed and installed into all current IntelliFill i.v. devices be
Quantity28 devices
DistributionNationwide Distribution including the states of CA, CO, CT, FL, IL, LA, MA, MD, MI, NC, NE, NY, OH, OR, PA, TN, TX, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.