Recall Z-0993-2014 ยท IPL Wheelchair, Standup

The Standing Company

Superstand stand-up wheelchair HPS-2 A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.

DateFeb 19, 2014 (initiated May 16, 13)
StatusTerminated
ReasonPower Superstand Standing Wheel chair, flex shaft (the flexible wand which supports the joystick control module) may fracture if it is bent repeatedly and /or beyond 90 degrees. A newly designed Flexshaft which corrects this problem is available at no charge.
Root causeComponent design/selection
ActionThe Standing Company sent an Urgent Medical Device Field Correction/Part Recall dated February 6, 2014, to all affected customers. The Urgent Recall notification identified the Flexshaft holding the control joystick on all Power Standing Wheelchairs were being replaced after investigation of complaints found the flexshaft could break because the inner spring of the shaft had failed, although the outer core remained intact. The notice explained how the failure would appear and that the break could happen without warning. The notice explained that the break would not cause the flexshaft to f
Quantityapproximately 119
DistributionNationwide Distribution including AL, CA, DE, IA, IL , IN, KS, KY, MI, MO, NJ, OH, OK, TX, WA, WI,and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032101POWERSTAND STAND-UP WHEELCHAIR
The Standing Co., Inc., Jan 22, 2004