Recall Z-0603-2014 ยท MNJ Valve, Pressure Relief, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Custom Cardiovascular Procedure Kits containing an Overpressure Safety Valve, LN130B, in the Arterial Recirculation Line

DateDec 27, 2013 (initiated Jun 4, 2013)
StatusTerminated
ReasonOver-Pressure Safety Valves not cleared with an indication for use on the arterial side of the extracorporeal bypass circuit.
Root causeComponent design/selection
ActionTerumo Cardiovascular issued Customer notification letters on June 4, 2013. Accounts advised to: Review this notice and decide to continue to use or not use the Cardiovascular Procedure Kits as configured until a replacement kit with a one-way valve indicated for use on the arterial side of the perfusion circuit is available. If you have questions, contact Terumo CVS Customer Service: 800-521-2818.
Quantity24,227.
DistributionNationwide Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K820297WM. HARVEY OVERPRESSURE SAFETY VALVE
C.R. Bard, Inc., Mar 12, 1982