Recall Z-1829-2014 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc.

UniCel DxH 600 Coulter Cellular Analysis System, Part No. B23858. Used for in vitro diagnostic use in screening patient populations found in clinical laboratories.

DateJun 20, 2014 (initiated Nov 8, 2013)
StatusTerminated
ReasonBeckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell differential and the reticulocyte.
Root causeUnder Investigation by firm
ActionAn Urgent Product Correction letter, dated 11/08/2013, was sent to all the customers who purchased the UniCel DxH 800 and UniCel DxH 600 Coulter Cellular Analysis Systems. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Beckman Coulter's Customer Support Center via website, http://www.beckmancoulter.com/customersupport/support, via phone at (800) 526-7694 in US and Canada. For outside of the US and Canada, cus
Quantity164 units total (106 units in US)
DistributionWorldwide Distribution -- US, Canada, Australia, China, Belgium, Brunei Darssalam, Colombia, Croatia, Czech Republic, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Korea, Kuwait, Macao, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Norway, Oman, Philipp
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081930UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Beckman Coulter, Inc., Dec 19, 2008