Recall Z-0818-2014 ยท MKW Cyclosporine

Siemens Healthcare Diagnostics, Inc.

Dimension Vista Cyclosporine Flex reagent cartridge an in vitro diagnostic test for the quantitative measurement of cyclosporine A (CSA) in human whole blood on the Dimension Vista System. Measurements are used as an aid in the management of heart, liver, and kidney transplant patients.

DateJan 23, 2014 (initiated Dec 5, 2013)
StatusTerminated
ReasonComplaints were received regarding variability in recovery of QC and patient results with certain lots of the Dimension Vista CSA Flex reagent cartridges.
Root causeNonconforming Material/Component
ActionSiegmens sent an Urgent Medical Device Recall lletter dated December 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use and discard any remaining inventory of the affected product. Customers were also instructed to complete the attached form and fax it to 302-631-8467. Customers were asked to retain the letter with their laboratory records, and forward the letter to any who have received the affected product. Customers with questions were instructed to contact Siemens Cust
Quantity1586
DistributionWorldwide Distribution - USA (nationwide) and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062236DIMENSION VISTA CYCLOSPORINE (CSA) FLEX REAGENT CARTRIDGE
Dade Behring, Inc., Aug 21, 2006