Recall Z-0714-2014 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Draeger Medical Systems, Inc.

Draeger Infinity Acute Care System Monitoring Solution

DateJan 10, 2014 (initiated Dec 10, 2013)
StatusTerminated
ReasonAfter a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
Root causeSoftware design
ActionAn Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.
Quantity439
DistributionUS Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113798INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
Draeger Medical Systems, Inc., Mar 22, 2012