Recall Z-0570-2014 ยท MKJ Automated External Defibrillators (Non-Wearable)
Philips Medical Systems, Inc.
Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the orde
| Date | Dec 24, 2013 (initiated Dec 10, 2013) |
| Status | Terminated |
| Reason | MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg |
| Root cause | Software design |
| Action | Philips Healthcare issued an Customer Information letter dated February 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer with the addendum to the Instructions for Use. If you have a ? in the parameter block and no CO2 waveform on the display, the waveform source is invalid. If invalid, check patient, confirm airway status and examine the FilterLine to make sure it is connected to the device and not kinked or pinched. For more information on the EtCO2 Option see Monitoring Carbon Dioxide chapter in the HeartSta |
| Quantity | 18,529 units |
| Distribution | US Nationwide and OUS. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K051134 | HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A Philips Medical Systems, Sep 8, 2005 |