Recall Z-0713-2014 ยท CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Radiometer America Inc

pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).

DateJan 10, 2014 (initiated Jun 10, 2013)
StatusTerminated
ReasonRADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
Root causeMixed-up of materials/components
ActionRadiometer sent a Product Notification letter dated June 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory and remove any affected product, discard all affected product in their possession, and complete page 2 of the letter and return to their Radiometer representative. Radiometer will replace any quantity of affected product they have at their location. Customers with questions were instructed to contact Radiometer America Technicial Support at 1-800-736-0600,
Quantity209 boxes/6 membranes
DistributionWorldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K895181ABL500 BLOOD GAS SYSTEM
Radiometer America, Inc., Oct 6, 1989
K922922ABL50 BLOOD GAS SYSTEM
Radiometer America, Inc., Aug 28, 1992
K935655ABL5 BLOOD GAS SYSTEM
Radiometer America, Inc., May 27, 1994