Recall Z-0630-2014 ยท KCY Tourniquet, Pneumatic

Stryker Instruments Div. of Stryker Corporation

Stryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder, Single Port, QuickConnect Rx only, STERILE

DateJan 6, 2014 (initiated Dec 6, 2013)
StatusTerminated
ReasonThe Color Cuff Sterile Disposable Tourniquet Cuffs with Quick-Connect Fitting may have an incorrect end-connector. A customer reported having received a luer connector instead of a quick connect fitting. An incorrect connector may lead to an inoperable system due to incompatibility with the Smart Pump.
Root causeComponent change control
ActionDistributor was notified by E-mail 11/26/2013, with follow-up UPS letters sent to Hospital Risk/ Materials Managers identifying the product, reason for the recall, and risk. The notice asked consignees to find and quarantine all implicated product, with hospitals requested to complete the Business Reply Form and FAX to 866-521-2762. Stryker will email a pre-paid shipper and credit all accounts. Question and concerns should be addressed to Kelly Jo Whipple 269-389-2921 or kellyjo.whipple@stryker.com
Quantity5 boxes (50 units)
DistributionUS Distribution: including states of: MI, OH, and TX.