Recall Z-0605-2014 ยท FRC Indicator, Biological Sterilization Process

Steris Corporation

VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.

DateDec 28, 2013 (initiated Sep 10, 2013)
StatusTerminated
ReasonSTERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.
Root causeUse error
ActionOn September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.
Quantity1,238 boxes
DistributionWorldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBI