Recall Z-0805-2014 ยท LNX

Sorin Group Italia S.r.l.

ORCHESTRA/ORCHESTRA PLUS Programmer

DateJan 22, 2014 (initiated Nov 20, 2013)
StatusTerminated
ReasonSorin has voluntarily issued a notification to physicians related to the overestimation of the residual longevity displayed by its programmer (ORCHESTRA or ORCHESTRA PLUS) during a follow-up exam of patients implanted with a REPLY or ESPRIT pacemaker.
Root causeSoftware design
ActionOn Nov 20 and Nov 25, 2013, the Dear Doctor Letter was sent by certified mail to 1,310 physicians who are following 13,003 patients implanted with a REPLU or ESPRIT pacemakers in the US. Sorin provided the following recommendations to the physicians: You should consider checking the battery impedance of the last follow-up exam. In case the battery impedance is greater than or equal to 3.5k, a follow-up visit must be scheduled within a maximum of 6 months from the last follow-up visit. When pacemaker operation is checked by the simple application of a magnet, a magnet rate less than 95 min-1
Quantity1,718
DistributionNationwide.