Recall Z-1929-2014 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Microgenics Corporation

MAS Omni Immune Controls

DateJun 26, 2014 (initiated Dec 2, 2013)
StatusTerminated
ReasonThe constituent, Inhibin A, was removed from MAS Omni IMMUNE and MAS Omni IMMUNE PRO controls; however the package insert was not updated to reflect the change.
Root causeLabeling Change Control
ActionUrgent Medical Device Correction letters were sent on December 2, 2013 to all affected customers and distributors via trackable mail.
Quantity3185
DistributionWorldwide Distribution-USA (nationwide) and the countries of Australia, China, Germany, Hong Kong and Malaysia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110616THERMO SCIENTIFIC MAS OMNI-IMMUNE, MAS OMNI IMMUNE PRO
Microgenics Corp., Jun 29, 2011