Recall Z-0640-2014 ยท GDZ Handle, Scalpel
Ansell Sandel Medical Industries LLC
Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Long Disposable Safety Handle with #15 Blade Manufactured for: Ansell Sandel Medical Solutions LLC 19736 Dearborn St. Chatsworth, CA 91311 USA Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Made in China Packaged in Mexico Ansell 2215-L Weighted Safety Scalpel, #3L Handle (8.75) with #15 Blade, Yellow Shield with TIME OU
| Date | Jan 8, 2014 (initiated Nov 12, 2013) |
| Status | Terminated |
| Reason | Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety shield which could compromise the integrity of the package and the sterility of the scalpel leading to injury. |
| Root cause | Under Investigation by firm |
| Action | The firm, Ansell, sent an "URGENT: VOLUNTARY RECALL OF ANSELL SANDEL WEIGHTED SAFET SCALPEL" letter dated November 11, 2013 to its customers and posted a website notification on November 11, 2013. The letter described the product, problem and actions to be taken. The customers were instructed: Do not ship SANDEL Weighted Safety Scalpel products 2210-L, 2210-N, 2211-L, 2211-N, 2215-L, 2215-N, or 2220-N to a hospital or end user with the listed lot numbers; quarantine SANDEL Weighted Safety Scalpel products in your inventory; and complete the attached response form and return it to attention: An |
| Quantity | 96 cases (1920 units) |
| Distribution | Worldwide Distribution: US (nationwide) including states of: AL, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MS, NC, NE, NY, OK, PA, PR, RI, TN, TX and WI; and country of: Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.